*** MIDAS logfile (05SEPpl1.2) *** Sa, 12 Aug 2006 13:50:22 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 12 Aug 2006 13:50:37 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 060812 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 349.25 stddev = 2.17 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_060812 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1714.6 1715.1 855.0 0.11 0.12 2.01 2 4.34 3419.0 3418.2 1692.5 -0.15 -0.16 2.02 3 8.67 6839.2 6840.1 3470.1 -0.08 -0.07 1.97 4 13.00 10258.0 10259.1 5111.2 -0.09 -0.08 2.01 5 19.51 15390.9 15390.7 7536.3 -0.06 -0.06 2.04 6 26.01 20529.9 20529.5 9997.5 -0.01 -0.01 2.05 7 32.51 25669.8 25672.0 12448.6 0.02 0.03 2.06 8 43.34 34252.9 34253.4 16270.3 0.10 0.10 2.11 9 54.17 42800.9 42802.1 20146.0 0.07 0.07 2.12 10 65.01 51364.4 51365.4 23758.1 0.07 0.07 2.16 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3704 to 110937 e- Average count rate measured: 790 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.16 +/- 0.02 e-/ADU RON = 4.68 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.001 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.12/-0.15 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay