*** MIDAS logfile (05SEPpl1.2) *** Fr, 11 Aug 2006 16:59:38 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Fr, 11 Aug 2006 16:59:53 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 060811 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 349.31 stddev = 2.18 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_060811 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.55 2013.2 2014.8 1015.8 0.04 0.05 1.98 2 5.11 4026.2 4026.8 2018.2 -0.03 -0.06 1.99 3 10.21 8048.8 8048.2 3978.8 -0.12 -0.14 2.02 4 15.31 12086.7 12086.3 5911.9 -0.02 -0.03 2.04 5 22.96 18129.5 18130.4 8868.7 -0.03 -0.03 2.04 6 30.62 24183.0 24184.0 11687.4 0.01 0.01 2.07 7 38.27 30243.2 30244.0 14410.3 0.05 0.06 2.10 8 51.02 40335.6 40338.2 19234.7 0.08 0.09 2.10 9 63.78 50389.5 50395.0 23480.4 0.02 0.03 2.15 10 76.53 60454.7 60456.3 27627.8 -0.00 0.00 2.19 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 4390 to 131790 e- Average count rate measured: 790 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.18 +/- 0.02 e-/ADU RON = 4.75 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.006 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.09/-0.13 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay