*** MIDAS logfile (05SEPpl1.2) *** Th, 10 Aug 2006 16:37:02 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Th, 10 Aug 2006 16:37:16 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 060810 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 349.18 stddev = 2.18 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_060810 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.54 2000.1 1999.9 997.5 0.07 0.09 2.01 2 5.07 3992.3 3991.9 1973.5 -0.12 -0.12 2.02 3 10.13 7994.9 7993.7 3957.8 -0.01 -0.02 2.02 4 15.19 11981.6 11977.2 5905.4 -0.09 -0.13 2.03 5 22.79 17973.5 17985.9 8824.8 -0.09 -0.03 2.04 6 30.38 23986.1 23984.5 11560.0 0.00 -0.01 2.07 7 37.98 30017.7 30014.8 14267.0 0.12 0.10 2.10 8 50.64 39985.8 40005.8 18784.0 0.02 0.06 2.13 9 63.29 50032.0 50028.9 23159.1 0.12 0.11 2.16 10 75.95 59949.5 59946.7 27114.8 -0.03 -0.04 2.21 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 4409 to 132159 e- Average count rate measured: 790 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.20 +/- 0.02 e-/ADU RON = 4.80 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.005 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.12/-0.13 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay