*** MIDAS logfile (05SEPpl1.2) *** We, 02 Aug 2006 17:21:27 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 02 Aug 2006 17:21:41 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 060802 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 349.45 stddev = 2.18 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_060802 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1581.7 1581.5 783.1 0.13 0.11 2.02 2 6.16 3150.9 3151.5 1577.1 -0.21 -0.20 2.00 3 12.31 6309.9 6309.8 3202.1 -0.06 -0.07 1.97 4 18.47 9459.7 9460.2 4704.8 -0.11 -0.10 2.01 5 27.70 14197.6 14197.4 6987.8 -0.05 -0.04 2.03 6 36.93 18935.3 18934.3 9145.7 -0.02 -0.02 2.07 7 46.16 23683.2 23680.8 11419.2 0.04 0.04 2.07 8 61.54 31586.5 31583.8 15292.4 0.07 0.07 2.07 9 76.93 39510.0 39503.4 18730.2 0.14 0.13 2.11 10 92.31 47376.0 47373.3 21855.2 0.06 0.07 2.17 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3403 to 101918 e- Average count rate measured: 513 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.15 +/- 0.02 e-/ADU RON = 4.70 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.001 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.14/-0.20 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay