*** MIDAS logfile (05SEPpl1.2) *** Su, 02 Jul 2006 14:21:07 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Su, 02 Jul 2006 14:21:22 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 060702 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 348.90 stddev = 2.00 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_060702 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1577.9 1578.5 782.0 0.12 0.12 2.02 2 6.16 3147.0 3147.3 1568.5 -0.15 -0.16 2.01 3 12.31 6299.0 6299.0 3135.9 -0.07 -0.07 2.01 4 18.47 9445.8 9446.0 4725.4 -0.10 -0.10 2.00 5 27.70 14175.6 14173.8 6996.2 -0.05 -0.06 2.03 6 36.93 18906.2 18905.8 9211.9 -0.02 -0.02 2.05 7 46.16 23644.4 23643.5 11473.8 0.03 0.03 2.06 8 61.54 31538.8 31539.0 15214.7 0.07 0.08 2.07 9 76.93 39436.6 39434.7 18723.1 0.10 0.10 2.11 10 92.31 47310.1 47307.2 22123.4 0.07 0.07 2.14 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3366 to 100906 e- Average count rate measured: 512 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.13 +/- 0.02 e-/ADU RON = 4.27 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.003 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.12/-0.16 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay