*** MIDAS logfile (05SEPpl1.2) *** Sa, 13 May 2006 15:05:12 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 13 May 2006 15:05:26 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 060513 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 349.00 stddev = 2.00 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_060513 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1611.0 1611.6 796.9 0.13 0.14 2.02 2 6.16 3209.7 3209.6 1612.9 -0.22 -0.23 1.99 3 12.31 6429.0 6428.6 3183.5 -0.05 -0.05 2.02 4 18.47 9637.5 9635.4 4747.5 -0.11 -0.12 2.03 5 27.70 14464.6 14462.8 7086.0 -0.04 -0.04 2.04 6 36.93 19291.3 19289.1 9315.1 -0.02 -0.01 2.07 7 46.16 24131.5 24128.6 11606.3 0.06 0.06 2.08 8 61.54 32171.9 32169.0 15397.3 0.05 0.05 2.09 9 76.93 40265.2 40258.3 19024.0 0.17 0.17 2.12 10 92.31 48248.2 48243.3 22500.5 0.03 0.03 2.14 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3451 to 103320 e- Average count rate measured: 522 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.14 +/- 0.01 e-/ADU RON = 4.28 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.001 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.17/-0.22 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay