*** MIDAS logfile (05SEPpl1.2) *** Sa, 06 May 2006 15:59:38 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 06 May 2006 15:59:52 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 060506 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 348.52 stddev = 2.01 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_060506 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1636.1 1636.1 814.4 0.11 0.11 2.01 2 6.16 3261.3 3261.1 1650.4 -0.18 -0.18 1.98 3 12.31 6528.7 6528.0 3267.3 -0.07 -0.07 2.00 4 18.47 9789.1 9788.2 4843.1 -0.10 -0.10 2.02 5 27.70 14690.6 14691.3 7193.5 -0.05 -0.04 2.04 6 36.93 19594.4 19593.3 9584.1 -0.01 -0.01 2.04 7 46.16 24504.2 24502.7 11823.6 0.04 0.04 2.07 8 61.54 32682.9 32680.6 15659.7 0.07 0.07 2.09 9 76.93 40877.8 40873.3 19178.9 0.13 0.13 2.13 10 92.31 49018.2 49014.3 23116.3 0.06 0.06 2.12 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3492 to 104603 e- Average count rate measured: 531 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.13 +/- 0.01 e-/ADU RON = 4.29 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.001 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.13/-0.18 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay