*** MIDAS logfile (05SEPpl1.2) *** Fr, 14 Apr 2006 16:50:42 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Fr, 14 Apr 2006 16:50:56 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 060414 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 348.37 stddev = 2.00 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_060414 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1643.4 1642.6 827.7 0.10 0.09 1.99 2 6.16 3276.8 3276.7 1671.8 -0.16 -0.15 1.96 3 12.31 6556.9 6557.5 3261.3 -0.09 -0.08 2.01 4 18.47 9833.5 9833.5 4860.5 -0.10 -0.10 2.02 5 27.70 14756.9 14756.5 7158.5 -0.05 -0.05 2.06 6 36.93 19682.8 19684.2 9556.8 -0.01 -0.01 2.06 7 46.16 24614.9 24613.3 11837.1 0.03 0.03 2.08 8 61.54 32835.2 32833.5 15671.0 0.08 0.08 2.10 9 76.93 41056.8 41052.5 19314.4 0.12 0.10 2.13 10 92.31 49249.1 49248.9 23159.2 0.08 0.07 2.13 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3511 to 105236 e- Average count rate measured: 533 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.14 +/- 0.01 e-/ADU RON = 4.27 +/- 0.02 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.002 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.11/-0.15 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay