*** MIDAS logfile (05SEPpl1.2) *** Fr, 07 Apr 2006 14:13:34 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Fr, 07 Apr 2006 14:13:48 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 060407 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 348.23 stddev = 2.00 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_060407 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1655.7 1655.8 839.7 0.11 0.11 1.97 2 6.16 3302.5 3302.4 1653.0 -0.15 -0.15 2.00 3 12.31 6608.2 6608.2 3289.2 -0.08 -0.08 2.01 4 18.47 9910.1 9910.3 4852.7 -0.10 -0.10 2.04 5 27.70 14872.5 14871.8 7260.5 -0.05 -0.05 2.05 6 36.93 19836.8 19836.3 9598.2 -0.01 -0.01 2.07 7 46.16 24806.9 24806.1 11900.0 0.03 0.03 2.08 8 61.54 33091.1 33091.6 15757.3 0.08 0.08 2.10 9 76.93 41369.5 41369.3 19527.1 0.09 0.10 2.12 10 92.31 49632.1 49631.0 23353.8 0.07 0.07 2.13 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3532 to 105883 e- Average count rate measured: 537 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.13 +/- 0.01 e-/ADU RON = 4.26 +/- 0.02 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.002 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.11/-0.15 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay