*** MIDAS logfile (05SEPpl1.2) *** Sa, 11 Mar 2006 19:26:00 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 11 Mar 2006 19:26:14 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 060311 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 350.27 stddev = 2.02 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_060311 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1670.6 1670.6 848.4 0.10 0.10 1.97 2 6.16 3332.4 3332.4 1651.6 -0.15 -0.15 2.02 3 12.31 6669.2 6669.3 3233.1 -0.08 -0.07 2.06 4 18.47 10001.1 10000.7 4944.9 -0.10 -0.10 2.02 5 27.70 15009.0 15008.0 7297.4 -0.05 -0.05 2.06 6 36.93 20018.2 20017.2 9761.5 -0.01 -0.02 2.05 7 46.16 25035.4 25034.3 12173.7 0.04 0.03 2.06 8 61.54 33392.4 33394.3 15672.8 0.07 0.08 2.13 9 76.93 41759.6 41754.7 19737.1 0.12 0.11 2.12 10 92.31 50089.9 50093.4 22833.0 0.07 0.08 2.19 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3633 to 108933 e- Average count rate measured: 542 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.17 +/- 0.03 e-/ADU RON = 4.40 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.002 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.11/-0.15 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay