*** MIDAS logfile (05SEPpl1.2) *** Fr, 03 Mar 2006 16:22:55 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Fr, 03 Mar 2006 16:23:09 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 060303 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 350.76 stddev = 2.03 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_060303 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1665.4 1665.0 831.8 0.14 0.15 2.00 2 6.16 3315.7 3316.1 1656.5 -0.26 -0.23 2.00 3 12.31 6645.1 6644.3 3284.4 -0.03 -0.04 2.02 4 18.47 9956.7 9957.0 4927.5 -0.13 -0.12 2.02 5 27.70 14947.1 14945.0 7247.7 -0.04 -0.05 2.06 6 36.93 19934.2 19932.8 9667.9 -0.02 -0.02 2.06 7 46.16 24939.6 24936.5 12064.8 0.07 0.07 2.07 8 61.54 33241.5 33241.3 15893.8 0.04 0.05 2.09 9 76.93 41619.9 41610.3 19763.3 0.21 0.19 2.11 10 92.31 49848.8 49843.1 23506.3 0.02 0.01 2.12 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3535 to 105827 e- Average count rate measured: 540 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.12 +/- 0.01 e-/ADU RON = 4.32 +/- 0.02 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.001 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.20/-0.25 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay