*** MIDAS logfile (05SEPpl1.2) *** Sa, 25 Feb 2006 13:02:24 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 25 Feb 2006 13:02:38 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 060225 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 350.52 stddev = 2.03 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_060225 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1677.9 1678.0 853.6 0.10 0.10 1.97 2 6.16 3346.4 3346.8 1672.1 -0.16 -0.15 2.00 3 12.31 6697.8 6697.7 3358.3 -0.07 -0.08 1.99 4 18.47 10044.3 10044.4 4991.1 -0.09 -0.09 2.01 5 27.70 15072.0 15072.6 7329.6 -0.05 -0.05 2.06 6 36.93 20103.5 20104.3 10102.1 -0.01 -0.01 1.99 7 46.16 25141.5 25141.5 12147.6 0.04 0.03 2.07 8 61.54 33538.3 33536.3 16165.8 0.09 0.08 2.07 9 76.93 41929.4 41929.7 19434.8 0.10 0.10 2.16 10 92.31 50298.1 50296.7 23239.9 0.07 0.06 2.16 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3637 to 109012 e- Average count rate measured: 545 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.17 +/- 0.03 e-/ADU RON = 4.40 +/- 0.06 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.002 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.10/-0.16 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay