*** MIDAS logfile (05SEPpl1.2) *** Fr, 17 Feb 2006 14:57:17 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Fr, 17 Feb 2006 14:57:31 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 060217 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 350.50 stddev = 2.03 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_060217 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1692.1 1692.4 871.9 0.11 0.11 1.94 2 6.16 3374.7 3375.2 1662.8 -0.15 -0.15 2.03 3 12.31 6753.4 6753.0 3332.4 -0.09 -0.09 2.03 4 18.47 10129.2 10129.8 4992.4 -0.09 -0.09 2.03 5 27.70 15199.0 15199.5 7439.5 -0.05 -0.05 2.04 6 36.93 20273.6 20274.5 9817.9 -0.01 -0.01 2.06 7 46.16 25351.7 25351.5 12238.1 0.03 0.02 2.07 8 61.54 33821.6 33823.0 16509.4 0.09 0.09 2.05 9 76.93 42276.2 42277.3 19681.7 0.09 0.09 2.15 10 92.31 50729.0 50728.1 23499.5 0.08 0.08 2.16 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3644 to 109242 e- Average count rate measured: 549 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.15 +/- 0.02 e-/ADU RON = 4.38 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.002 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.11/-0.15 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay