*** MIDAS logfile (05SEPpl1.2) *** Sa, 11 Feb 2006 14:02:08 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 11 Feb 2006 14:02:28 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 060211 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 350.51 stddev = 2.03 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_060211 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1685.8 1685.4 857.4 0.10 0.12 1.97 2 6.16 3362.3 3361.8 1689.8 -0.15 -0.15 1.99 3 12.31 6728.9 6729.6 3306.8 -0.08 -0.06 2.03 4 18.47 10090.4 10090.7 4996.6 -0.10 -0.10 2.02 5 27.70 15143.1 15143.6 7431.7 -0.05 -0.04 2.04 6 36.93 20199.2 20198.0 9791.7 -0.00 -0.01 2.06 7 46.16 25258.5 25259.3 12237.1 0.03 0.03 2.06 8 61.54 33693.0 33692.4 16061.9 0.08 0.07 2.10 9 76.93 42126.0 42120.9 19888.8 0.10 0.09 2.12 10 92.31 50532.0 50530.3 23465.4 0.06 0.06 2.15 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3623 to 108601 e- Average count rate measured: 547 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.15 +/- 0.02 e-/ADU RON = 4.35 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.003 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.11/-0.15 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay