*** MIDAS logfile (05SEPpl1.2) *** Su, 05 Feb 2006 16:35:05 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Su, 05 Feb 2006 16:35:19 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 060205 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 350.62 stddev = 2.04 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_060205 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1696.9 1696.9 873.3 0.09 0.11 1.94 2 6.16 3384.3 3383.9 1784.9 -0.15 -0.15 1.90 3 12.31 6771.7 6771.2 3349.5 -0.08 -0.08 2.02 4 18.47 10155.0 10156.1 4946.1 -0.10 -0.09 2.05 5 27.70 15239.9 15240.3 7599.0 -0.04 -0.04 2.01 6 36.93 20327.2 20328.1 10018.9 -0.00 -0.00 2.03 7 46.16 25420.0 25418.8 12198.2 0.04 0.03 2.08 8 61.54 33907.6 33908.2 16127.3 0.08 0.08 2.10 9 76.93 42389.6 42388.5 20116.5 0.10 0.09 2.11 10 92.31 50851.7 50851.2 23254.0 0.06 0.06 2.19 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3682 to 110346 e- Average count rate measured: 551 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.17 +/- 0.03 e-/ADU RON = 4.42 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.002 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.11/-0.15 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay