*** MIDAS logfile (05SEPpl1.2) *** Fr, 03 Feb 2006 18:14:40 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Fr, 03 Feb 2006 18:14:54 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 060203 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 350.53 stddev = 2.03 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_060203 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1694.3 1693.8 864.4 0.07 0.07 1.96 2 6.16 3385.5 3385.8 1714.2 -0.05 -0.04 1.97 3 12.31 6760.6 6760.8 3338.6 -0.23 -0.22 2.02 4 18.47 10160.0 10159.4 5019.4 -0.05 -0.05 2.02 5 27.69 15247.4 15247.2 7467.5 0.00 0.00 2.04 6 36.93 20336.7 20337.1 10034.3 0.03 0.04 2.03 7 46.16 25435.2 25432.2 12331.6 0.08 0.08 2.06 8 61.54 33924.9 33925.6 16308.2 0.12 0.12 2.08 9 76.93 42322.3 42319.4 19898.4 -0.08 -0.09 2.13 10 92.31 50877.0 50877.5 23431.8 0.09 0.10 2.17 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3660 to 109916 e- Average count rate measured: 551 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.16 +/- 0.02 e-/ADU RON = 4.39 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.006 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.12/-0.23 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay