*** MIDAS logfile (05SEPpl1.2) *** Fr, 27 Jan 2006 13:11:39 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Fr, 27 Jan 2006 13:11:53 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 060127 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 350.58 stddev = 2.03 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_060127 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1687.9 1688.8 860.9 0.10 0.11 1.96 2 6.16 3367.3 3367.5 1667.9 -0.15 -0.16 2.02 3 12.31 6739.4 6738.5 3337.4 -0.07 -0.09 2.02 4 18.47 10106.6 10106.7 5044.0 -0.10 -0.10 2.00 5 27.70 15166.8 15167.1 7430.9 -0.05 -0.04 2.04 6 36.93 20230.8 20231.3 9738.2 -0.01 -0.00 2.08 7 46.16 25299.2 25299.2 12148.7 0.04 0.04 2.08 8 61.54 33746.8 33745.6 16065.6 0.08 0.08 2.10 9 76.93 42190.6 42192.0 19690.3 0.10 0.10 2.14 10 92.31 50613.5 50615.6 23423.5 0.07 0.07 2.16 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3654 to 109551 e- Average count rate measured: 548 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.16 +/- 0.02 e-/ADU RON = 4.39 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.003 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.10/-0.16 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay