*** MIDAS logfile (05SEPpl1.2) *** Sa, 14 Jan 2006 15:22:41 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 14 Jan 2006 15:22:55 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 060114 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 350.63 stddev = 2.03 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_060114 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1698.9 1698.8 862.7 0.10 0.11 1.97 2 6.16 3388.8 3388.9 1682.0 -0.16 -0.15 2.01 3 12.31 6783.1 6782.2 3359.6 -0.08 -0.09 2.02 4 18.47 10173.0 10172.0 5024.7 -0.09 -0.10 2.02 5 27.70 15266.2 15265.5 7492.0 -0.05 -0.04 2.04 6 36.93 20363.6 20361.9 9820.4 -0.00 -0.01 2.07 7 46.16 25463.8 25462.7 12299.6 0.03 0.03 2.07 8 61.54 33968.9 33968.1 16159.8 0.09 0.09 2.10 9 76.93 42462.6 42464.4 19931.4 0.09 0.10 2.13 10 92.31 50945.9 50943.9 23474.2 0.07 0.07 2.17 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3677 to 110263 e- Average count rate measured: 552 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.16 +/- 0.02 e-/ADU RON = 4.39 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.003 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.10/-0.15 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay