*** MIDAS logfile (05SEPpl1.2) *** We, 11 Jan 2006 15:45:27 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 11 Jan 2006 15:45:40 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 060111 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 350.59 stddev = 2.03 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_060111 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1685.9 1685.3 849.6 0.09 0.11 1.98 2 6.16 3364.8 3365.7 1682.6 -0.13 -0.11 2.00 3 12.31 6733.2 6733.6 3319.2 -0.09 -0.10 2.03 4 18.47 10099.8 10100.4 4995.5 -0.08 -0.11 2.02 5 27.70 15153.2 15155.1 7509.0 -0.06 -0.09 2.02 6 36.93 20215.2 20216.8 9724.2 -0.01 -0.04 2.08 7 46.16 25276.0 25276.9 12229.6 0.02 -0.02 2.07 8 61.54 33727.3 33728.2 16071.1 0.10 0.05 2.10 9 76.93 42142.4 42149.5 19746.3 0.06 0.03 2.13 10 92.31 50592.1 50712.1 23947.1 0.10 0.29 2.11 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3595 to 108035 e- Average count rate measured: 548 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.13 +/- 0.01 e-/ADU RON = 4.33 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.006 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.19/-0.12 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay