*** MIDAS logfile (05SEPpl1.2) *** Mo, 26 Dec 2005 14:42:12 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Mo, 26 Dec 2005 14:42:26 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 051226 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 350.75 stddev = 2.03 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_051226 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1680.8 1681.3 844.8 0.10 0.11 1.99 2 6.16 3355.2 3355.1 1696.2 -0.12 -0.11 1.98 3 12.31 6711.3 6711.1 3309.7 -0.10 -0.11 2.03 4 18.47 10068.6 10067.8 4986.4 -0.09 -0.10 2.02 5 27.70 15107.0 15107.2 7440.3 -0.06 -0.06 2.03 6 36.93 20152.6 20152.4 9697.2 -0.01 -0.01 2.08 7 46.16 25196.9 25198.9 12059.1 0.01 0.02 2.09 8 61.54 33622.7 33626.4 16031.8 0.09 0.10 2.10 9 76.93 42019.1 42018.8 19588.0 0.07 0.07 2.15 10 92.31 50440.6 50442.7 23538.6 0.10 0.11 2.14 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3622 to 108668 e- Average count rate measured: 546 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.15 +/- 0.01 e-/ADU RON = 4.38 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.004 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.11/-0.12 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay