*** MIDAS logfile (03SEPpl1.2) *** Fr, 09 Dec 2005 18:37:37 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Fr, 09 Dec 2005 18:37:51 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 051209 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 351.11 stddev = 2.04 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_051209 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1671.4 1671.2 826.2 0.09 0.08 2.02 2 6.16 3337.1 3337.2 1651.5 -0.11 -0.11 2.02 3 12.31 6676.8 6676.8 3282.2 -0.09 -0.09 2.03 4 18.47 10016.4 10016.3 4900.1 -0.08 -0.08 2.04 5 27.70 15028.4 15028.3 7507.8 -0.06 -0.06 2.00 6 36.93 20047.3 20047.9 9704.6 -0.01 -0.01 2.07 7 46.16 25064.9 25065.6 12207.3 0.01 0.01 2.05 8 61.54 33448.4 33448.2 15748.3 0.10 0.09 2.12 9 76.93 41794.2 41795.6 19538.1 0.05 0.06 2.14 10 92.31 50177.8 50181.5 23226.5 0.10 0.11 2.16 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3616 to 108574 e- Average count rate measured: 543 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.16 +/- 0.02 e-/ADU RON = 4.42 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.005 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.11/-0.11 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay