*** MIDAS logfile (05SEPpl1.1) *** Sa, 26 Nov 2005 16:03:56 Page 1 *** MIDAS logfile (vers. 05SEPpl1.1) *** Sa, 26 Nov 2005 16:04:11 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 051126 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 351.17 stddev = 2.04 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_051126 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1670.6 1671.1 846.7 0.11 0.10 1.97 2 6.16 3333.3 3332.8 1678.0 -0.13 -0.15 1.99 3 12.31 6669.0 6669.2 3286.4 -0.09 -0.09 2.03 4 18.47 10001.7 10003.0 4971.9 -0.11 -0.10 2.01 5 27.70 15009.3 15010.8 7241.1 -0.06 -0.05 2.07 6 36.93 20021.3 20022.5 9742.8 -0.01 -0.01 2.05 7 46.16 25036.7 25036.4 12005.5 0.02 0.03 2.09 8 61.54 33401.5 33403.7 15732.5 0.08 0.09 2.12 9 76.93 41756.6 41752.4 19804.9 0.10 0.09 2.11 10 92.31 50105.4 50106.7 23154.4 0.09 0.10 2.16 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3604 to 108064 e- Average count rate measured: 542 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.16 +/- 0.02 e-/ADU RON = 4.39 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.003 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.10/-0.14 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay