*** MIDAS logfile (03SEPpl1.2) *** Fr, 04 Nov 2005 15:05:59 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Fr, 04 Nov 2005 15:06:15 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 051104 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 350.85 stddev = 2.04 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_051104 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1669.4 1669.4 848.2 0.13 0.12 1.97 2 6.16 3325.8 3326.5 1666.1 -0.22 -0.21 2.00 3 12.31 6661.6 6661.4 3345.5 -0.06 -0.06 1.99 4 18.47 9987.2 9985.7 4922.1 -0.11 -0.11 2.03 5 27.70 14989.7 14988.2 7321.5 -0.04 -0.04 2.05 6 36.93 19992.8 19991.7 9786.9 -0.00 -0.00 2.04 7 46.16 25005.7 25003.3 12026.1 0.06 0.05 2.08 8 61.54 33339.2 33339.6 15881.6 0.05 0.06 2.10 9 76.93 41715.9 41712.3 19452.1 0.16 0.15 2.14 10 92.31 49997.8 49995.9 22982.5 0.03 0.03 2.18 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3629 to 108679 e- Average count rate measured: 541 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.17 +/- 0.02 e-/ADU RON = 4.43 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.001 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.15/-0.22 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay