*** MIDAS logfile (03SEPpl1.2) *** Sa, 29 Oct 2005 16:54:43 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Sa, 29 Oct 2005 16:54:59 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 051029 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 350.92 stddev = 2.03 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_051029 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1676.3 1676.3 851.6 0.12 0.12 1.97 2 6.16 3343.9 3344.2 1689.2 -0.15 -0.15 1.98 3 12.31 6692.6 6693.0 3367.6 -0.09 -0.09 1.99 4 18.47 10038.2 10038.6 4935.6 -0.10 -0.09 2.03 5 27.70 15063.0 15063.6 7456.3 -0.05 -0.05 2.02 6 36.93 20093.7 20093.9 9751.7 -0.01 -0.01 2.06 7 46.16 25126.5 25126.6 12206.3 0.03 0.03 2.06 8 61.54 33521.2 33521.2 15981.8 0.09 0.08 2.10 9 76.93 41904.3 41904.4 19592.8 0.09 0.09 2.14 10 92.31 50278.7 50281.0 23013.6 0.08 0.08 2.18 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3649 to 109467 e- Average count rate measured: 544 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.18 +/- 0.02 e-/ADU RON = 4.42 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.004 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.12/-0.15 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay