*** MIDAS logfile (03SEPpl1.2) *** Sa, 08 Oct 2005 12:30:50 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Sa, 08 Oct 2005 12:31:05 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 051008 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 350.90 stddev = 2.03 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_051008 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1685.8 1686.0 854.9 0.13 0.12 1.97 2 6.16 3359.9 3360.3 1682.1 -0.20 -0.20 2.00 3 12.31 6729.1 6728.7 3330.8 -0.05 -0.05 2.02 4 18.47 10086.6 10086.9 4994.8 -0.12 -0.11 2.02 5 27.70 15140.4 15140.0 7484.3 -0.04 -0.04 2.02 6 36.93 20192.8 20192.3 9905.0 -0.01 -0.01 2.04 7 46.16 25258.4 25256.7 12176.5 0.06 0.06 2.07 8 61.54 33678.7 33675.6 16099.1 0.06 0.06 2.09 9 76.93 42136.5 42133.3 19609.6 0.15 0.15 2.15 10 92.31 50500.3 50494.1 23497.4 0.03 0.02 2.15 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3638 to 108979 e- Average count rate measured: 547 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.16 +/- 0.02 e-/ADU RON = 4.38 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.002 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.15/-0.20 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay