*** MIDAS logfile (03SEPpl1.2) *** We, 21 Sep 2005 15:26:47 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** We, 21 Sep 2005 15:27:03 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 050921 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 350.85 stddev = 2.03 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_050921 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1681.4 1681.1 843.5 0.12 0.11 1.99 2 6.16 3351.6 3351.4 1667.8 -0.20 -0.19 2.01 3 12.31 6711.4 6711.1 3282.0 -0.06 -0.06 2.04 4 18.47 10061.0 10061.7 4950.8 -0.12 -0.10 2.03 5 27.70 15103.1 15101.1 7508.7 -0.04 -0.04 2.01 6 36.93 20142.2 20141.2 9735.9 -0.01 -0.01 2.07 7 46.16 25192.0 25190.3 12255.0 0.05 0.05 2.06 8 61.54 33595.5 33592.2 16163.8 0.07 0.06 2.08 9 76.93 42024.7 42021.2 19648.9 0.14 0.14 2.14 10 92.31 50381.5 50380.8 23458.0 0.04 0.05 2.15 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3612 to 108239 e- Average count rate measured: 546 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.15 +/- 0.02 e-/ADU RON = 4.35 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.002 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.14/-0.19 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay