*** MIDAS logfile (03SEPpl1.2) *** Th, 25 Aug 2005 18:14:10 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Th, 25 Aug 2005 18:14:26 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 050825 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 350.87 stddev = 2.03 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_050825 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1694.9 1694.6 858.0 0.12 0.11 1.98 2 6.16 3379.6 3379.2 1685.2 -0.16 -0.18 2.01 3 12.31 6763.9 6764.1 3328.5 -0.09 -0.08 2.03 4 18.47 10143.9 10143.9 5083.6 -0.10 -0.09 2.00 5 27.70 15222.8 15221.9 7474.6 -0.05 -0.05 2.04 6 36.93 20304.9 20304.2 9890.2 -0.01 -0.01 2.05 7 46.16 25393.1 25391.3 12475.7 0.04 0.04 2.04 8 61.54 33871.2 33870.7 16108.8 0.08 0.09 2.10 9 76.93 42354.9 42347.6 19797.7 0.12 0.11 2.14 10 92.31 50796.9 50796.9 23505.9 0.06 0.07 2.16 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3664 to 109808 e- Average count rate measured: 550 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.16 +/- 0.02 e-/ADU RON = 4.39 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.002 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.11/-0.17 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay