*** MIDAS logfile (03SEPpl1.2) *** Su, 21 Aug 2005 21:51:46 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Su, 21 Aug 2005 21:52:02 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 050821 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 350.97 stddev = 2.02 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_050821 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1692.0 1692.1 860.7 0.09 0.10 1.97 2 6.16 3377.5 3377.4 1704.4 -0.12 -0.12 1.98 3 12.31 6756.1 6757.0 3351.6 -0.10 -0.10 2.02 4 18.47 10136.2 10136.0 5012.8 -0.09 -0.09 2.02 5 27.70 15208.4 15209.3 7418.6 -0.06 -0.05 2.05 6 36.93 20287.7 20288.2 9887.7 -0.01 -0.01 2.05 7 46.16 25368.9 25366.8 12491.5 0.03 0.01 2.03 8 61.54 33848.1 33849.8 16139.3 0.09 0.10 2.10 9 76.93 42301.4 42301.5 20063.9 0.07 0.07 2.11 10 92.31 50773.9 50776.1 23246.8 0.10 0.10 2.18 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3659 to 109798 e- Average count rate measured: 550 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.16 +/- 0.03 e-/ADU RON = 4.37 +/- 0.06 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.004 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.10/-0.12 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay