*** MIDAS logfile (03SEPpl1.2) *** Th, 11 Aug 2005 14:30:07 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Th, 11 Aug 2005 14:30:23 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 050811 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 350.84 stddev = 2.02 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_050811 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1688.8 1688.9 864.0 0.12 0.12 1.95 2 6.16 3366.5 3366.4 1669.8 -0.18 -0.18 2.02 3 12.31 6739.4 6739.3 3348.5 -0.08 -0.08 2.01 4 18.47 10105.4 10106.4 4984.3 -0.11 -0.10 2.03 5 27.70 15166.9 15166.6 7406.8 -0.05 -0.05 2.05 6 36.93 20229.5 20229.8 9818.0 -0.01 -0.01 2.06 7 46.16 25300.9 25299.9 12151.1 0.04 0.04 2.08 8 61.54 33746.4 33744.9 15954.1 0.08 0.08 2.12 9 76.93 42202.1 42198.8 19781.4 0.13 0.12 2.13 10 92.31 50608.2 50607.8 23428.0 0.06 0.06 2.16 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3652 to 109432 e- Average count rate measured: 548 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.16 +/- 0.02 e-/ADU RON = 4.38 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.002 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.12/-0.18 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay