*** MIDAS logfile (03SEPpl1.2) *** We, 27 Jul 2005 14:44:09 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** We, 27 Jul 2005 14:44:24 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 050727 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 350.91 stddev = 2.02 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_050727 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1692.2 1692.2 855.1 0.10 0.11 1.98 2 6.16 3374.1 3373.9 1705.0 -0.17 -0.18 1.98 3 12.31 6754.3 6753.8 3315.2 -0.07 -0.08 2.04 4 18.47 10127.9 10128.6 4993.2 -0.10 -0.09 2.03 5 27.70 15199.7 15197.6 7530.7 -0.04 -0.05 2.02 6 36.93 20273.6 20274.0 9869.4 -0.01 -0.00 2.05 7 46.16 25352.3 25353.4 12256.0 0.04 0.04 2.07 8 61.54 33818.3 33819.4 16203.1 0.08 0.09 2.09 9 76.93 42283.3 42280.9 19966.6 0.11 0.10 2.12 10 92.31 50720.3 50719.6 23048.4 0.07 0.07 2.20 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3681 to 110326 e- Average count rate measured: 549 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.18 +/- 0.03 e-/ADU RON = 4.40 +/- 0.06 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.002 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.11/-0.17 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay