*** MIDAS logfile (03SEPpl1.2) *** Th, 23 Jun 2005 15:36:10 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Th, 23 Jun 2005 15:36:26 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 050623 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 350.42 stddev = 2.01 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_050623 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1721.6 1721.6 877.2 0.15 0.15 1.96 2 6.16 3428.2 3428.6 1702.1 -0.24 -0.22 2.01 3 12.31 6867.9 6866.6 3423.4 -0.05 -0.05 2.01 4 18.47 10292.8 10291.7 5090.6 -0.12 -0.12 2.02 5 27.70 15450.7 15449.4 7559.0 -0.04 -0.04 2.04 6 36.93 20606.8 20605.0 10045.4 -0.01 -0.01 2.05 7 46.16 25777.7 25775.0 12555.8 0.07 0.07 2.05 8 61.54 34366.0 34361.1 16456.2 0.06 0.05 2.09 9 76.93 43005.5 42999.9 20245.1 0.17 0.17 2.12 10 92.31 51522.0 51516.0 23628.6 0.01 0.01 2.18 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3728 to 111551 e- Average count rate measured: 558 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.17 +/- 0.02 e-/ADU RON = 4.35 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.000 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.17/-0.23 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay