*** MIDAS logfile (03SEPpl1.2) *** Sa, 18 Jun 2005 13:20:37 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Sa, 18 Jun 2005 13:20:53 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 050618 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 350.32 stddev = 2.01 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_050618 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1742.5 1742.3 873.2 0.10 0.10 2.00 2 6.16 3475.0 3475.7 1762.4 -0.16 -0.15 1.97 3 12.31 6954.4 6954.4 3455.3 -0.09 -0.09 2.01 4 18.47 10430.8 10430.5 5213.8 -0.09 -0.10 2.00 5 27.70 15651.8 15652.2 7648.0 -0.05 -0.05 2.05 6 36.93 20879.2 20879.6 10252.5 0.00 -0.00 2.04 7 46.16 26106.9 26109.3 12402.6 0.03 0.04 2.10 8 61.54 34830.0 34830.2 16610.7 0.09 0.09 2.10 9 76.93 43534.8 43535.0 20444.4 0.09 0.08 2.13 10 92.31 52232.1 52233.6 24284.8 0.07 0.07 2.15 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3755 to 112578 e- Average count rate measured: 565 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.16 +/- 0.02 e-/ADU RON = 4.34 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.002 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.10/-0.15 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay