*** MIDAS logfile (03SEPpl1.2) *** Tu, 14 Jun 2005 08:21:13 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Tu, 14 Jun 2005 08:21:29 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 050614 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 350.00 stddev = 2.01 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_050614 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1725.6 1725.6 868.7 0.13 0.12 1.99 2 6.16 3439.8 3440.1 1705.2 -0.17 -0.17 2.02 3 12.31 6885.4 6885.7 3454.9 -0.08 -0.07 1.99 4 18.47 10325.4 10325.6 5109.4 -0.10 -0.10 2.02 5 27.70 15495.3 15495.5 7523.4 -0.05 -0.05 2.06 6 36.93 20667.5 20667.6 10026.7 -0.01 -0.01 2.06 7 46.16 25848.6 25847.0 12450.6 0.04 0.04 2.08 8 61.54 34476.4 34477.5 16364.1 0.08 0.08 2.11 9 76.93 43109.3 43106.2 20291.6 0.11 0.10 2.12 10 92.31 51703.1 51702.8 23963.2 0.06 0.06 2.16 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3720 to 111474 e- Average count rate measured: 560 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.16 +/- 0.02 e-/ADU RON = 4.33 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.002 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.12/-0.17 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay