*** MIDAS logfile (03SEPpl1.2) *** Fr, 03 Jun 2005 14:31:13 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Fr, 03 Jun 2005 14:31:29 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 050603 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 351.84 stddev = 2.06 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_050603 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1721.9 1722.0 884.4 0.10 0.11 1.95 2 6.16 3434.6 3435.2 1680.7 -0.16 -0.14 2.04 3 12.31 6874.3 6873.6 3379.9 -0.08 -0.09 2.03 4 18.47 10310.2 10308.9 5100.6 -0.09 -0.10 2.02 5 27.70 15471.3 15470.8 7619.9 -0.05 -0.05 2.03 6 36.93 20635.8 20636.5 9932.9 -0.01 -0.01 2.08 7 46.16 25807.4 25806.3 12515.4 0.04 0.03 2.06 8 61.54 34425.7 34426.4 16276.1 0.08 0.09 2.12 9 76.93 43040.3 43037.9 20246.2 0.10 0.10 2.13 10 92.31 51629.1 51630.5 23792.9 0.06 0.07 2.17 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3726 to 111716 e- Average count rate measured: 559 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.16 +/- 0.02 e-/ADU RON = 4.46 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.003 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.10/-0.15 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay