*** MIDAS logfile (03SEPpl1.2) *** Su, 29 May 2005 17:32:11 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Su, 29 May 2005 17:32:27 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 050529 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 351.95 stddev = 2.06 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_050529 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1728.9 1728.6 866.1 0.12 0.10 2.00 2 6.16 3446.1 3447.0 1735.7 -0.20 -0.18 1.99 3 12.31 6901.4 6901.0 3412.3 -0.06 -0.06 2.02 4 18.47 10347.6 10346.7 5117.5 -0.10 -0.10 2.02 5 27.70 15529.9 15528.5 7665.3 -0.04 -0.04 2.03 6 36.93 20713.8 20712.0 10196.9 -0.00 -0.01 2.03 7 46.16 25907.2 25905.6 12540.1 0.06 0.05 2.07 8 61.54 34546.6 34547.1 16570.5 0.07 0.07 2.08 9 76.93 43210.4 43204.6 20281.1 0.13 0.12 2.13 10 92.31 51804.3 51803.3 23774.7 0.04 0.04 2.18 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3746 to 112266 e- Average count rate measured: 561 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.17 +/- 0.02 e-/ADU RON = 4.46 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.002 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.13/-0.19 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay