*** MIDAS logfile (03SEPpl1.2) *** Su, 22 May 2005 17:38:41 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Su, 22 May 2005 17:38:57 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 050522 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 352.01 stddev = 2.08 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_050522 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1734.7 1734.8 859.0 0.12 0.12 2.02 2 6.16 3458.5 3458.3 1730.6 -0.18 -0.18 2.00 3 12.31 6923.6 6922.4 3405.8 -0.07 -0.08 2.03 4 18.47 10381.6 10381.8 5173.0 -0.10 -0.10 2.01 5 27.70 15580.6 15579.6 7569.7 -0.04 -0.04 2.06 6 36.93 20781.6 20781.8 10121.8 -0.01 -0.00 2.05 7 46.16 25990.8 25989.4 12589.4 0.05 0.05 2.06 8 61.54 34664.7 34662.7 16570.6 0.08 0.08 2.09 9 76.93 43344.9 43342.2 20301.8 0.11 0.11 2.14 10 92.31 51984.5 51982.4 24167.4 0.05 0.05 2.15 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3733 to 111859 e- Average count rate measured: 563 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.15 +/- 0.02 e-/ADU RON = 4.49 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.002 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.12/-0.18 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay