*** MIDAS logfile (03SEPpl1.2) *** Su, 08 May 2005 16:09:39 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Su, 08 May 2005 16:09:55 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 050508 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 351.89 stddev = 2.07 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_050508 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1745.1 1745.2 873.3 0.13 0.11 2.00 2 6.16 3478.9 3479.2 1745.9 -0.18 -0.17 1.99 3 12.31 6964.4 6964.8 3498.6 -0.07 -0.07 1.99 4 18.47 10443.5 10444.4 5218.6 -0.10 -0.09 2.00 5 27.70 15673.4 15673.5 7783.5 -0.04 -0.04 2.01 6 36.93 20906.1 20906.1 10187.6 -0.00 -0.00 2.05 7 46.16 26145.4 26145.9 12606.6 0.05 0.05 2.07 8 61.54 34871.6 34869.5 16555.5 0.08 0.07 2.11 9 76.93 43603.2 43601.6 20618.3 0.11 0.11 2.11 10 92.31 52288.4 52285.3 24275.0 0.04 0.04 2.15 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3758 to 112601 e- Average count rate measured: 566 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.15 +/- 0.02 e-/ADU RON = 4.47 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.002 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.12/-0.17 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay