*** MIDAS logfile (03SEPpl1.2) *** Sa, 30 Apr 2005 15:49:46 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Sa, 30 Apr 2005 15:50:02 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 050430 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 351.71 stddev = 2.06 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_050430 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1750.3 1750.1 869.3 0.13 0.13 2.01 2 6.16 3487.3 3487.0 1752.8 -0.21 -0.21 1.99 3 12.31 6985.1 6983.5 3499.3 -0.04 -0.06 2.00 4 18.47 10469.6 10469.2 5111.4 -0.11 -0.11 2.05 5 27.70 15715.1 15713.4 7671.7 -0.04 -0.04 2.05 6 36.93 20959.6 20957.7 10262.7 -0.01 -0.01 2.04 7 46.16 26217.3 26213.6 12455.5 0.06 0.06 2.10 8 61.54 34954.5 34951.6 16723.6 0.06 0.06 2.09 9 76.93 43734.2 43728.6 20561.0 0.15 0.15 2.13 10 92.31 52409.3 52404.7 24234.6 0.02 0.02 2.16 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3776 to 113072 e- Average count rate measured: 568 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.16 +/- 0.02 e-/ADU RON = 4.45 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.001 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.15/-0.21 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay