*** MIDAS logfile (03SEPpl1.2) *** Su, 24 Apr 2005 17:37:24 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Su, 24 Apr 2005 17:37:40 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 050424 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 351.81 stddev = 2.06 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_050424 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1745.7 1745.1 882.1 0.09 0.09 1.98 2 6.16 3484.4 3485.1 1759.2 -0.11 -0.09 1.98 3 12.31 6969.3 6969.7 3450.8 -0.11 -0.11 2.02 4 18.47 10457.5 10458.7 5176.3 -0.08 -0.08 2.02 5 27.70 15689.8 15690.6 7721.2 -0.05 -0.06 2.03 6 36.93 20930.4 20931.8 10170.5 -0.00 -0.01 2.06 7 46.16 26168.0 26169.7 12735.6 0.02 0.01 2.05 8 61.54 34922.4 34924.3 16687.4 0.11 0.10 2.09 9 76.93 43622.6 43627.7 20579.6 0.04 0.04 2.12 10 92.31 52380.6 52382.9 24367.0 0.10 0.09 2.15 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3748 to 112478 e- Average count rate measured: 567 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.15 +/- 0.02 e-/ADU RON = 4.41 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.005 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.10/-0.11 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay