*** MIDAS logfile (03SEPpl1.2) *** Sa, 16 Apr 2005 17:25:29 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Sa, 16 Apr 2005 17:25:45 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 050416 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 351.77 stddev = 2.06 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_050416 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1734.1 1733.4 873.6 0.12 0.11 1.98 2 6.16 3456.5 3456.0 1719.5 -0.19 -0.18 2.01 3 12.31 6920.0 6919.5 3418.2 -0.07 -0.07 2.02 4 18.47 10376.2 10375.8 5123.8 -0.10 -0.10 2.03 5 27.70 15573.2 15572.1 7561.2 -0.04 -0.04 2.06 6 36.93 20771.2 20769.5 10054.3 -0.00 -0.00 2.07 7 46.16 25976.1 25974.6 12606.7 0.04 0.05 2.06 8 61.54 34645.5 34643.3 16753.3 0.08 0.08 2.07 9 76.93 43320.0 43317.7 20569.3 0.11 0.11 2.11 10 92.31 51951.1 51949.1 23971.7 0.04 0.05 2.17 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3720 to 111474 e- Average count rate measured: 563 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.15 +/- 0.02 e-/ADU RON = 4.42 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.002 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.12/-0.18 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay