*** MIDAS logfile (03SEPpl1.2) *** Su, 10 Apr 2005 22:34:16 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Su, 10 Apr 2005 22:34:31 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 050410 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 351.51 stddev = 2.06 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_050410 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1759.0 1759.3 867.0 0.09 0.07 2.03 2 6.16 3511.6 3511.9 1752.9 -0.10 -0.11 2.00 3 12.31 7021.9 7023.7 3483.3 -0.12 -0.11 2.02 4 18.47 10538.0 10539.2 5232.7 -0.07 -0.07 2.01 5 27.70 15810.8 15812.0 7667.9 -0.05 -0.05 2.06 6 36.93 21091.4 21092.1 10143.2 0.00 -0.00 2.08 7 46.16 26368.4 26369.1 12691.5 0.01 0.01 2.08 8 61.54 35189.8 35191.1 16770.5 0.11 0.11 2.10 9 76.93 43956.1 43960.3 20789.8 0.03 0.04 2.11 10 92.31 52779.2 52783.3 24497.0 0.09 0.10 2.15 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3779 to 113380 e- Average count rate measured: 571 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.15 +/- 0.01 e-/ADU RON = 4.43 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.004 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.11/-0.11 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay