*** MIDAS logfile (03SEPpl1.2) *** Sa, 09 Apr 2005 18:15:26 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Sa, 09 Apr 2005 18:15:41 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 050409 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 351.80 stddev = 2.06 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_050409 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1746.1 1746.3 866.2 0.10 0.11 2.02 2 6.16 3483.2 3483.3 1741.5 -0.14 -0.15 2.00 3 12.31 6969.3 6970.2 3440.5 -0.09 -0.08 2.03 4 18.47 10453.6 10452.9 5185.9 -0.09 -0.10 2.02 5 27.70 15686.0 15685.3 7860.1 -0.05 -0.05 2.00 6 36.93 20923.9 20924.5 10182.4 -0.00 -0.00 2.05 7 46.16 26165.4 26166.5 12745.6 0.04 0.04 2.05 8 61.54 34903.6 34903.5 16619.4 0.09 0.08 2.10 9 76.93 43631.1 43629.7 20236.2 0.09 0.09 2.16 10 92.31 52345.4 52344.9 24053.6 0.07 0.06 2.18 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3802 to 113983 e- Average count rate measured: 567 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.18 +/- 0.03 e-/ADU RON = 4.49 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.002 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.10/-0.15 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay