*** MIDAS logfile (03SEPpl1.2) *** Su, 13 Mar 2005 15:41:30 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Su, 13 Mar 2005 15:41:45 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 050313 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 351.04 stddev = 2.03 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_050313 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1776.5 1775.9 894.5 0.10 0.10 1.99 2 6.16 3541.9 3541.9 1763.6 -0.16 -0.14 2.01 3 12.31 7087.9 7088.5 3467.4 -0.08 -0.07 2.04 4 18.47 10630.3 10630.8 5254.2 -0.09 -0.08 2.02 5 27.70 15952.7 15952.6 7868.8 -0.04 -0.04 2.03 6 36.93 21280.0 21278.8 10354.2 0.01 0.00 2.06 7 46.16 26610.9 26610.4 12959.9 0.05 0.05 2.05 8 61.54 35497.4 35495.9 16912.8 0.10 0.09 2.10 9 76.93 44376.4 44372.8 21076.2 0.11 0.10 2.11 10 92.31 53191.2 53185.1 24693.8 -0.00 -0.02 2.15 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3808 to 114017 e- Average count rate measured: 576 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.14 +/- 0.02 e-/ADU RON = 4.35 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.002 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.10/-0.15 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay