*** MIDAS logfile (03SEPpl1.2) *** Su, 06 Mar 2005 15:20:58 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Su, 06 Mar 2005 15:21:15 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 050306 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 350.72 stddev = 2.03 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_050306 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1808.8 1808.8 906.9 0.09 0.09 1.99 2 6.16 3609.2 3609.3 1799.7 -0.14 -0.14 2.01 3 12.31 7222.5 7222.8 3597.9 -0.08 -0.08 2.01 4 18.47 10833.4 10833.6 5315.1 -0.08 -0.08 2.04 5 27.70 16257.3 16256.8 7945.1 -0.04 -0.04 2.05 6 36.93 21685.7 21686.3 10581.4 0.01 0.01 2.05 7 46.16 27115.2 27116.3 13206.0 0.04 0.04 2.05 8 61.54 36174.0 36172.2 17093.7 0.09 0.09 2.12 9 76.93 45207.7 45209.1 21259.1 0.07 0.07 2.13 10 92.31 54235.4 54233.8 24891.2 0.04 0.04 2.18 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3924 to 117644 e- Average count rate measured: 587 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.17 +/- 0.02 e-/ADU RON = 4.41 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.003 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.09/-0.14 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay