*** MIDAS logfile (03SEPpl1.2) *** Mo, 28 Feb 2005 15:44:48 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Mo, 28 Feb 2005 15:45:04 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 050228 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 355.62 stddev = 2.00 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_050228 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1630.7 1630.1 749.6 0.12 0.13 2.18 2 6.16 3253.7 3253.3 1529.3 -0.17 -0.17 2.13 3 12.31 6513.5 6513.0 3079.1 -0.12 -0.12 2.12 4 18.47 9769.8 9769.9 4693.1 -0.13 -0.13 2.08 5 27.70 14663.5 14662.7 7097.6 -0.08 -0.08 2.07 6 36.93 19560.4 19561.3 9458.6 -0.03 -0.03 2.07 7 46.16 24464.4 24462.5 11820.7 0.02 0.01 2.07 8 61.54 32644.2 32645.9 15503.4 0.09 0.10 2.11 9 76.93 40822.1 40823.8 19162.9 0.13 0.13 2.13 10 92.31 49000.3 49001.6 23038.8 0.16 0.16 2.13 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3466 to 104185 e- Average count rate measured: 530 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.13 +/- 0.01 e-/ADU RON = 4.26 +/- 0.02 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.009 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.16/-0.17 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay