*** MIDAS logfile (03SEPpl1.2) *** Sa, 12 Feb 2005 16:42:51 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Sa, 12 Feb 2005 16:43:08 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 050212 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 350.75 stddev = 2.04 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_050212 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1640.6 1640.5 844.2 0.08 0.10 1.94 2 6.16 3273.6 3273.6 1662.4 -0.11 -0.12 1.97 3 12.31 6546.7 6547.2 3277.3 -0.11 -0.10 2.00 4 18.47 9821.1 9821.8 4916.8 -0.09 -0.09 2.00 5 27.70 14735.1 14734.8 7308.4 -0.06 -0.06 2.02 6 36.93 19656.7 19656.7 9555.2 -0.01 -0.01 2.06 7 46.16 24576.2 24575.4 11848.2 0.02 0.01 2.07 8 61.54 32793.9 32796.7 15613.3 0.10 0.10 2.10 9 76.93 40980.1 40982.2 19375.7 0.07 0.06 2.12 10 92.31 49196.9 49201.0 22867.5 0.11 0.11 2.15 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3531 to 105898 e- Average count rate measured: 532 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.15 +/- 0.02 e-/ADU RON = 4.38 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.002 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.11/-0.12 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay