*** MIDAS logfile (03SEPpl1.2) *** Sa, 22 Jan 2005 22:11:27 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Sa, 22 Jan 2005 22:11:43 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 050122 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 350.15 stddev = 2.07 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_050122 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1676.7 1677.0 879.8 0.09 0.10 1.91 2 6.16 3344.6 3343.7 1712.5 -0.14 -0.16 1.95 3 12.31 6691.3 6691.6 3319.2 -0.09 -0.08 2.02 4 18.47 10036.1 10036.8 5028.1 -0.09 -0.08 2.00 5 27.70 15059.6 15060.6 7467.4 -0.05 -0.04 2.02 6 36.93 20089.1 20090.0 9881.2 0.01 0.01 2.03 7 46.16 25117.3 25117.7 12145.5 0.03 0.03 2.07 8 61.54 33511.7 33509.8 16079.8 0.10 0.09 2.08 9 76.93 41881.4 41880.6 19645.5 0.08 0.07 2.13 10 92.31 50251.3 50250.3 23282.4 0.07 0.06 2.16 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3622 to 108529 e- Average count rate measured: 544 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.16 +/- 0.02 e-/ADU RON = 4.46 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.001 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.10/-0.15 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay