*** MIDAS logfile (03SEPpl1.2) *** Mo, 10 Jan 2005 13:52:58 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Mo, 10 Jan 2005 13:53:15 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 050110 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 350.04 stddev = 2.07 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_050110 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1673.7 1674.0 864.9 0.09 0.11 1.94 2 6.16 3339.4 3339.2 1714.9 -0.13 -0.12 1.95 3 12.31 6678.8 6679.5 3340.1 -0.10 -0.10 2.00 4 18.47 10018.3 10019.0 5024.4 -0.10 -0.10 1.99 5 27.70 15032.8 15033.9 7436.8 -0.05 -0.06 2.02 6 36.93 20054.0 20053.3 9831.2 -0.00 -0.01 2.04 7 46.16 25073.8 25075.3 12202.3 0.02 0.02 2.05 8 61.54 33455.2 33457.3 16060.8 0.10 0.09 2.08 9 76.93 41807.5 41809.4 19642.8 0.07 0.06 2.13 10 92.31 50184.1 50188.9 23160.3 0.10 0.10 2.17 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3621 to 108557 e- Average count rate measured: 543 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.16 +/- 0.02 e-/ADU RON = 4.47 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.002 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.10/-0.12 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay