*** MIDAS logfile (03SEPpl1.2) *** Su, 19 Dec 2004 13:15:52 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Su, 19 Dec 2004 13:16:08 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 041219 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 350.15 stddev = 2.07 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_041219 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1655.5 1656.4 848.0 0.12 0.13 1.95 2 6.16 3299.2 3299.9 1673.1 -0.19 -0.18 1.97 3 12.31 6604.4 6604.4 3310.4 -0.07 -0.07 2.00 4 18.47 9902.4 9901.3 4954.9 -0.10 -0.11 2.00 5 27.70 14861.3 14859.3 7374.8 -0.04 -0.05 2.02 6 36.93 19820.5 19819.8 9794.2 -0.01 -0.01 2.02 7 46.16 24789.4 24788.5 12060.5 0.04 0.05 2.06 8 61.54 33060.7 33059.9 15921.6 0.07 0.07 2.08 9 76.93 41351.1 41346.3 19739.4 0.13 0.13 2.09 10 92.31 49583.0 49582.3 22932.4 0.06 0.06 2.16 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3554 to 106407 e- Average count rate measured: 537 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.15 +/- 0.02 e-/ADU RON = 4.44 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.000 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.13/-0.18 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay