*** MIDAS logfile (03SEPpl1.2) *** Tu, 07 Dec 2004 15:25:37 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Tu, 07 Dec 2004 15:25:53 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 041207 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 350.05 stddev = 2.08 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_041207 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1658.5 1659.1 864.0 0.06 0.05 1.92 2 6.16 3314.5 3314.6 1682.7 -0.01 -0.02 1.97 3 12.31 6620.3 6620.7 3349.5 -0.12 -0.13 1.98 4 18.47 9932.2 9933.1 4985.6 -0.11 -0.10 1.99 5 27.70 14901.1 14901.6 7396.1 -0.08 -0.08 2.01 6 36.93 19878.0 19878.0 9911.8 -0.03 -0.03 2.01 7 46.16 24852.8 24853.8 12101.3 -0.01 -0.01 2.05 8 61.54 33166.0 33169.1 15928.1 0.08 0.09 2.08 9 76.93 41439.2 41441.8 19420.4 0.04 0.04 2.13 10 92.31 49803.8 49804.4 22914.6 0.19 0.19 2.17 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3600 to 108073 e- Average count rate measured: 539 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.17 +/- 0.03 e-/ADU RON = 4.50 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.002 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.19/-0.13 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay